| Wavelengths (nm)(a) | 1310 ± 20/1550 ± 20/1625 ± 15 |
| Dynamic range at 20 µs (dB)(b, c) | 1310/1550 model: dynamic range = 46/46 dB 1310/1550/1625 model: dynamic range = 45/45/45 dB |
| Event dead zone (m)(d) | 0.5 |
| Attenuation dead zone (m)(e) | 2.5 |
| Distance range (km) | 0.1 to 400 |
| Pulse width (ns) | 3 to 20 000 |
| Linearity (dB/dB)(a) | ±0.03 |
| Loss threshold (dB) | 0.01 |
| Loss resolution (dB) | 0.001 |
| Sampling resolution (m) | 0.04 to 10 |
| Sampling points | Up to 256 000 |
| Distance uncertainty (m)(f) | ±(0.75 + 0.0025 % × distance + sampling resolution) |
| Measurement time | User-defined (maximum: 60 minutes) |
| Typical real-time refresh (Hz) | 4 |
| Stable source output power (dBm)(g) | 1.5 |
| Reflectance (dB)(a) | ±2 |
| a. Typical. b. Typical dynamic range with a three-minute averaging at SNR = 1. c. 50 dB dynamic range can be obtained with a G.655 fiber with a 30 minutes averaging time. d. Typical for reflectance from –35 dB to –55 dB, at 3-ns pulse. e. Typical at 1310 nm, for reflectance at –55 dB. Attenuation dead zone is 3.5 m typical at 1310 nm with reflectance below –45 dB. f. Does not include uncertainty due to fiber index. g. Typical output power value at 1550 nm. All specifications valid at 23 °C ± 2 °C with an FC/APC connector, unless otherwise specified. | |
| Display | 8-inch (203 mm), 1280×800, color touchscreen (viewable in sunlight) |
| Interfaces | USB-A ports (2) USB-C port with power delivery RJ45 LAN 10/100/1000 Mbit/s |
| RF comms(a, b) | Bluetooth, Wi-Fi, 2G/3G/4G LTE, GNSS (GPS/GALILEO/QZSS) |
| Storage | >20,000 OTDR SOR traces |
| Battery | Rechargeable LiFePO4 battery, up to 10 hours(c) of operation as per Telcordia (Bellcore) GR-196-CORE |
| Power supply | Input: AC/DC adapter, 100 to 240 V AC, 50 to 60 Hz, 1.5 A max. Output: 5 to 20 V DC, 3.0 A max., 45 W max., USB-C power delivery standard supported |
| Weight (including battery and module) | 2.4 kg (5.3 lb) |
| Size (H × W × D) | 198 mm × 249 mm × 71 mm (7.8 in × 9.8 in × 2.8 in) |
| Temperature | Operating: –10 °C to 50 °C (14 °F to 122 °F) Storage: –40 °C to 70 °C (–40 °F to 158 °F) |
| Relative humidity | 0 % to 95 % non-condensing |
| Warranty (year) | 1 |
| Built-in power meter specifications (GeX) (optional)(d) | |
| Calibrated wavelengths (nm) | 850, 1300, 1310, 1342, 1358, 1490, 1550, 1577, 1625, 1650 |
| Selectable wavelengths (nm) | 850, 1300, 1310, 1342, 1358, 1490, 1550, 1577, 1625, 1650 |
| Power range (dBm)(e) | 27 to –50 |
| Uncertainty (%)(f) | ±5 % |
| Display resolution (dB) | 0.01 = max to –40 dBm 0.1 = –40 dBm to –50 dBm |
| Tone detection (Hz) | 270/330/1000/2000 |
| VFL specifications | VFL (optional) / High-power VFL (optional) |
| Operation mode | VFL (optional): Flashing (slow/fast) and continuous; High-power VFL (optional): Flashing (slow/fast) and continuous |
| Flashing frequency (Hz) | VFL (optional): 1 or 4; High-power VFL (optional): 1 or 4 |
| Wavelength (nm) (typical) | VFL (optional): 650; High-power VFL (optional): 660 |
| Emitter type | VFL (optional): Laser; High-power VFL (optional): Laser |
| Power output (mW) (max.) | VFL (optional): 1; High-power VFL (optional): 5 |
| Distance range (km) (typical)(g) | VFL (optional): 7; High-power VFL (optional): 12 |
| Laser safety class | VFL (optional): 2; High-power VFL (optional): 3R |
| Laser safety | Complies with FDA 1040.10 and IEC 60825-1:2014-05. Without VFL (option): IEC 60825-1:2014-05. With VFL (option): IEC 60825-1:2014-05, IEC/CEI 60825-1:2014. With high power VFL (option): IEC 60825-1:2014-05, IEC/CEI 60825-1:2014. Applicability: Class 1M, 2M and 3R. WARNING: Viewing the laser output with telescopic optical instruments (for example, telescopes and binoculars) may pose an eye hazard and thus the user should not direct the beam into an area where such instruments are likely to be used. |
| a. Free 36-month basic data plan provided. b. Restrictions may apply depending on country/region which will prevent EXFO from providing mobile connectivity. Contact EXFO for details. c. Battery life varies significantly based on device configuration, usage, network and feature configuration, signal strength, settings and other factors. d. At 23 ºC ± 1 ºC, 1550 nm and FC connector. With modules in idle mode. Battery operated after 30-minute warm-up. e. Typical. f. At calibration conditions. g. Depends on fiber attenuation and ambient light conditions. | |