Vollständige Spezifikationen

T200S - High-power continuously tunable laser

Übersicht
Faster and highly reliable testing for optical components, including photonic integrated circuits.
T200S - High-power continuously tunable laser

Leistungsmerkmale

  • Up to 200 nm/s scanning speed
  • 10 dBm over tunable spectral range
  • Excellent SSSER
  • Wavelength coverage: O and CL bands
  • Wavelength tuning and continuous sweep modes
  • Active mode-hop-free operation

Anwendungen

  • Passive optical components
  • Photonic integrated circuits
  • Multipurpose tunable laser for R&D

Spezifikationen

Technical specifications
Wavelength range (nm) 1260 - 1360 (T200S-O); 1500 - 1630 (T200S-CL)
Wavelength uncertainty a (pm) ±5 (typical); ±20
Wavelength repeatability b (pm) ±5 (typical)
Wavelength stability c (pm) ±5
Wavelength resolution setting (pm) 1
Maximum speed (nm/s) 100 (200 optional)
Adjustable speed (nm/s) 0.5, 1, 2, 5, 10, 20, 50, 100 (200 optional)
Mode-hop-free operation Active mode-hop cancellation
Minimum wavelength sweep trigger resolution j (pm) 0.5
Nominal output power over full wavelength range (dBm) 10
Power stability c (dB) ±0.01 (typical)
Linewidth d (10 µs integration time) (kHz) < 25 (typical)
Linewidth d (100 µs integration time) (kHz) < 250 (typical)
Side mode suppression ratio e (SMSR) (dB) > 50 (typical)
Signal-to-source spontaneous emission ratio f (SSSER) (dB) 90 (typical)
Signal-to-total-source spontaneous emission ratio g (STSSER) (dB) 75 (typical)
Relative intensity noise h (RIN) (dB/Hz) –150 (typical)
Optical fiber type i PM optical fiber, FC/APC connector
PER (dB) 17 (typical)
All specifications given at constant temperature ± 1 °C, after wavelength full reference, between 18 °C and 28 °C (unless otherwise stated), after a 60-minute warm-up and for wavelength not equal to water peak.
a. TUNE mode, high accuracy setting.
b. TUNE mode, high accuracy setting and for wavelength tuning from low to high wavelengths.
c. Over one hour and at a temperature of 21 °C ± 1 °C. Measured at 10 dBm. Stability expressed as ± half difference between max and min values measured within 60-minute window.
d. In TUNE mode, optimized linewidth setting, at 10 dBm and at a temperature of 21 °C ± 1 °C.
e. At 10 dBm, at 21 °C ± 1 °C. For T200S-O: 1270 nm to 1360 nm.
f. Measured over a bandwidth of 0.1 nm, at 10 dBm, for the center wavelength.
g. Measured over a span of 100 nm with an exclusion zone of ±0.6 nm around the signal. Laser at center wavelength.
h. RIN within 100 MHz - 3 GHz range with RBW = 30 kHz. For T200S-O: 1270 nm to 1360 nm.
i. Fiber’s slow axis and polarization aligned with key connector.
j. With output trigger PULSE activated, available for all scanning speeds. Triggers generated at regular wavelength intervals during the sweep; some restrictions apply to wavelength scanning span of the laser according to speed and resolution; not compatible with bidirectional scanning.
General specifications
Environmental conditions
Operating temperature 15 °C to 35 °C (59 °F to 95 °F)
Operating humidity < 80% (non-condensing)
Warm-up time (hour) 1
Physical footprint
Size (L x H x D) 217 mm x 173 mm x 441 mm (8 9/16 in x 6 13/16 in x 17 3/8 in)
Weight 9 kg (20 lb)
Connectivity
Monitor 7 in capacitive touchscreen
Remote communication Ethernet RJ45 LAN 10/100/1000 Mbit/s
Electrical BNC ports 1x trigger IN, 1x trigger OUT, power monitoring and wavelength monitoring
USB ports USB 3.0 (1), USB 2.0 (2)
Security
Laser safety Class 1M
Power supply 100 - 240 V AC; 50/60 Hz; 0.65 - 0.3 A
Accessories (sold separately) Rackmount 4U tablet accommodating 2 units